• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS TINA-QUANT HEMOGLOBIN A1CDX GEN. 3; GLYCOSYLATED HEMOGLOBIN ASSAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS TINA-QUANT HEMOGLOBIN A1CDX GEN. 3; GLYCOSYLATED HEMOGLOBIN ASSAY Back to Search Results
Catalog Number 07826915700
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/09/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The analyzer serial number is: (b)(6).The customer replaced the hemolyzing reagent.The investigation is ongoing.
 
Event or Problem Description
There was an allegation of questionable a1c-2 tina-quant hemoglobin a1c gen.2 results for 1 patient sample on a cobas pure c 303 analytical unit.The customer reported a positive bias on their proficiency results which prompted them to performed a patient comparison.The initial a1c-2 was 6.01%.The sample was repeated on a c11 analyzer and the result was 5.46%.
 
Additional Manufacturer Narrative
Based on the calibration and qc data, a general reagent issue could be excluded.The customer changing the hemolyzing reagent resolved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TINA-QUANT HEMOGLOBIN A1CDX GEN. 3
Common Device Name
GLYCOSYLATED HEMOGLOBIN ASSAY
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18608822
Report Number1823260-2024-00273
Device Sequence Number19519553
Product Code LCP
UDI-Device Identifier07613336157824
UDI-Public07613336157824
Combination Product (Y/N)Y
Initial Reporter CountryIZ
PMA/510(K) Number
K072714
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2024
Device Catalogue Number07826915700
Device Lot Number74419901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/09/2024
Supplement Date Received by Manufacturer02/28/2024
Initial Report FDA Received Date01/30/2024
Supplement Report FDA Received Date03/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-