• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. CORKSCREW FT, BC, SUTURETAPE, 4.75M; BIO SOFT TISSUE FIXATN FASTNR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. CORKSCREW FT, BC, SUTURETAPE, 4.75M; BIO SOFT TISSUE FIXATN FASTNR Back to Search Results
Model Number CORKSCREW FT, BC, SUTURETAPE, 4.75M
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 12/05/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
On 01/09/2024, it was reported by a facility representative via e-mail that an ar-1927bct-475 suture anchor pulled out.This occurred on (b)(6) 2023.No additional information was provided, and additional information has been asked.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to improper bone preparation and/or prying/leveraging during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORKSCREW FT, BC, SUTURETAPE, 4.75M
Common Device Name
BIO SOFT TISSUE FIXATN FASTNR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
emily shafer
8009337001
MDR Report Key18608892
Report Number1220246-2024-00616
Device Sequence Number9033469
Product Code MAI
UDI-Device Identifier00888867237858
UDI-Public00888867237858
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K173788
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 04/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCORKSCREW FT, BC, SUTURETAPE, 4.75M
Device Catalogue NumberAR-1927BCT-475
Device Lot Number15082121
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/09/2024
Supplement Date Received by Manufacturer04/02/2025
Initial Report FDA Received Date01/30/2024
Supplement Report FDA Received Date04/16/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/18/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-