MAUDE Adverse Event Report: SMITH & NEPHEW, INC. COCR 12/14 FEM HEAD 26 + 0; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. COCR 12/14 FEM HEAD 26 + 0; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 71302600
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 01/12/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, after thr surgery had been performed on (b)(6) 2020, the patient fell and suffered from dislocation.This adverse event was addressed by revision surgery on (b)(6) 2024, in which when trying manual reduction, it was found the cocr head was disassociated from tandem bipolar.Therefore, the cocr 12/14 fem head 26 + 0, tndm bp shl/xlpe lnr 40od 26id and sl-plus integr mia stem with ti/ha 2 were exchange.Patient's current health status is unknown.
 
Additional Manufacturer Narrative
H3, h6: the associated devices were returned and evaluated.The visual inspection revealed a few scratches on both devices.The devices show signs of normal wear.Dimensional evaluation performed on the device revealed that all metal components were found to be in tolerance per associated prints.The clinical/medical investigation concluded that, based on the information provided, the patients¿ fall cannot be ruled out as a contributing factor to patient¿s dislocation and subsequent revision.The impact to the patient beyond the reported fall, the dislocation, and the subsequent revision cannot be determined since the patient¿s current health status is unknown.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for total hip systems revealed that this has been identified as a that dislocation has been identified in warnings and precautions, that may result from improper selection of the neck length and cup, stem positioning, muscle looseness and/or malpositioning of components.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COCR 12/14 FEM HEAD 26 + 0
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18608931
Report Number1020279-2024-00253
Device Sequence Number13016014
Product Code LPH
UDI-Device Identifier03596010194077
UDI-Public03596010194077
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K963509
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number71302600
Device Lot Number20FM06297
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2024
Supplement Date Received by Manufacturer10/31/2024
Initial Report FDA Received Date01/30/2024
Supplement Report FDA Received Date11/01/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/09/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PN: 71325040, LOT: 14HM20538.; PN: 75000175, LOT: UNKNOWN.
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient SexFemale
-
-