SMITH & NEPHEW, INC. STEM IMPACTOR ROD; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
|
Back to Search Results |
|
| Catalog Number |
71364011 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/12/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a thr surgery, the threads of a stem impactor rod broke off the end of the pummel.The procedure was resumed, without any delay, using the same device.Patient was not harmed as consequence of this problem.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: h3, h6: the device was not returned for evaluation; therefore, a device analysis could not be performed.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history based on the historical data revealed a similar event for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A review of previously escalated cases concluded that this failure was further investigated and addressed through design change evaluations.This lot had been manufactured before such evaluation.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|