| Model Number |
PCF-H190I |
| Medical Device Problem Code |
Microbial Contamination of Device (2303)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/09/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The suspect device has been returned to olympus for evaluation.The physical device evaluation has been completed.Prior to the device evaluation, the device was sent out for additional microbiological testing.The hygiene microbiological investigation (hmi) report indicated the channels of the scope were cultured and 3 colony forming units cfu /endoscope of micrococcaceae and bacillaceae.The results obtained comply with the target level for an endoscope subjected to high level disinfection and rinsed with sterile water.The device evaluation found scope connector was broken and scratches on the suction cylinder and charged couple device cover lens.The customer completed the cleaning, disinfection, and sterilization (cds) checklist for endoscopes, but no answers were provided on the relevant information of hmi result at the customer site.There was no patient infection.There have been no deviations or deficiencies or concerns in reprocessing.The investigation is ongoing and follow up with the customer is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
| |
|
Event or Problem Description
|
|
It was reported that, the colonovideoscope tested positive for an unexpected contamination.The issue was found during routine microbiological culture of the scope during reprocessing.All channels were sampled.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with the instructions for use (ifu) before repair, the results conformed to the regulation's recommendation.The following is included in the device ifu: "an insufficiently reprocessed endoscope and/or accessory may pose an infection control risk to the patients and/or operators who touch them." olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|