| Catalog Number |
130003 |
| Medical Device Problem Codes |
Leak/Splash (1354); Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Operator of device is unknown; no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient circuit connector was too loose and wouldn't work correctly, causing internal leakage.There was no patient involvement and no patient harm/adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
D4.Primary udi number: (b)(4).Investigation summary: the affected device was returned for evaluation.Visual inspection performed.Device was in good condition, no signs of external damage.Connected to oxygen supply and performed lock off leak test.Customer's reported issue was confirmed.User interface/demand valve screw cap was found loose.Tightened and secured patient output connector and demand valve screw cap to correct the issue.The service history review identified there was no indication that the complaint was related to a service of the device within the review period.
|
| |
|