SMITHS MEDICAL ASD, INC. PNEUPAC VENTILATORS VENTIPAC; VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR) Back to Search Results
Catalog Number 130003
Medical Device Problem Codes Leak/Splash (1354); Loose or Intermittent Connection (1371)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Operator of device is unknown; no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event or Problem Description
It was reported that the patient circuit connector was too loose and wouldn't work correctly, causing internal leakage.There was no patient involvement and no patient harm/adverse event reported.
 
Additional Manufacturer Narrative
D4.Primary udi number: (b)(4).Investigation summary: the affected device was returned for evaluation.Visual inspection performed.Device was in good condition, no signs of external damage.Connected to oxygen supply and performed lock off leak test.Customer's reported issue was confirmed.User interface/demand valve screw cap was found loose.Tightened and secured patient output connector and demand valve screw cap to correct the issue.The service history review identified there was no indication that the complaint was related to a service of the device within the review period.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PNEUPAC VENTILATORS VENTIPAC
Common Device Name
VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
3350 granada ave n, suite 100
oakdale MN 55128
Manufacturer Contact
reed covert
6000 nathan lane n
minneapolis, MN 55442
2247062300
MDR Report Key18610466
Report Number2183161-2024-00119
Device Sequence Number2406058
Product Code BTL
UDI-Device Identifier10610586038259
UDI-Public(01)10610586038259(11)070511(21)0703299
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K030803
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 11/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number130003
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/08/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/05/2024
Supplement Date Received by Manufacturer11/04/2024
Initial Report FDA Received Date01/30/2024
Supplement Report FDA Received Date11/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/11/2007
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-