MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3545001 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
01/04/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient who underwent primary breast augmentation with 500cc mentor memorygel breast implants experienced bilateral rupture post procedure.The ruptures were confirmed during explant.Explant and replacement with mentor gel was performed on (b)(6), 2024.See 1645337-2024-01091 for contralateral prosthesis report.
|
| |
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: rupture mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The impacted product was received by the failure analysis lab on (b)(6) 2024.The investigation of the returned device was completed by the failure analysis lab on (b)(6) 2024.Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted visual inspection of the returned device.Visual analysis of the returned sample revealed that the breast implant had parallel lines of shell wear within a tear measuring approximately 1.2cm on anterior view.The evaluation determined that the possible cause of the rupture is consistent with normal wear.Shell wear suggests in-vivo folding or creasing of the device.This may be the result of the following factors: continuous and sustained stresses to the device such as too small breast pocket and folding or wrinkling of the shell in the breast pocket.In some cases, the breast implants may also wear out over time.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|