In addition to the phenomena identified in b5, the following phenomena were also identified during evaluation: there was a leak in the biopsy channel, and there were tears on the biopsy channel wall.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of the device history record and historical complaints analysis was conducted and no deviations that could have caused or contributed to the reported issue were identified.Based on the results of the investigation, and since the device malfunction was not confirmed during evaluation a definitive root cause could not be determined.The event can be prevented by following the instructions for use which state: chapter 3 preparation and inspection before each case, prepare and inspect this instrument as instructed below.Inspect other equipment to be used with this instrument as instructed in their respective instruction manuals.Should any irregularity be observed after inspection, follow the instructions as described in chapter 5, ¿troubleshooting¿.If this instrument malfunctions, do not use it.Return it to olympus for repair as described in section 5.3, ¿returning the endoscope for repair¿ on page 99.Warning: using an endoscope that is not functioning properly may compromise patient or operator safety and may result in more severe equipment damage.Chapter 5 troubleshooting if the endoscope is visibly damaged, does not function as expected, or is found to have irregularities during the inspection described in chapter 3, ¿preparation and inspection¿, do not use the endoscope.Contact olympus.Some problems that appear to be malfunctions may be correctable by referring to section 5.1, ¿troubleshooting guide¿.If the problem cannot be resolved by the described remedial action, stop using the endoscope and send it to olympus for repair.Warning: never use the endoscope on a patient if an irregularity is observed.Damage or an irregularity in the instrument may compromise patient or user safety and may result in more severe equipment damage.Should any irregularity in the function of the endoscope and/or endoscopic image be observed during use, stop the examination immediately and carefully withdraw the endoscope from the patient as described in section 5.2, ¿withdrawal of the endoscope with an irregularity¿ on page 97.Olympus will continue to monitor field performance for this device.
|