| Model Number |
9125E |
| Medical Device Problem Code |
Device Fell (4014)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided following the conclusion of the investigation.H3 other text : device not returned to the manufacturer.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2024 getinge became aware of an issue with 91e-series washer disinfector.A customer allegation was related to the door issue.As it was stated, load (a container) had fallen off the loading cart and obstructed the door.No getinge technician service visit was required.Machine returned to normal working condition.So far, we have not been informed about any adverse consequences of reported issue, however we decided to report the issue based on the potential as the load falling to the ground may result in a serious injury if the situation was to reoccur.
|
| |
|
Additional Manufacturer Narrative
|
|
On january 19th, 2024 getinge became aware of an issue with the 91e series washer disinfector, with the model name 9125e and serial number (b)(6).The washer disinfector was manufactured on 27th september, 2019.The reported issue is related to container falling from the cart to the floor and obstructed the door.Trend review of customer product complaints with the same issue involved on this type of devices reported within the last 5 years was performed but did not provide any signals that would warrant further scrutiny.Based on the investigation and information provided by the subject matter expert, the device was found to be working according to specification and the issue was resolved by hospital¿s employees.The root cause of the issue is most likely associated with incorrect use of the product - user error ¿ customer did not load the trolley correctly according to user manual.When the incident occurred, the product was directly involved.The device was not being used for treatment or diagnosis of the patient.There was no injury reported until date, nevertheless we decided to report the issue based on the potential as this kind of malfunction could lead to a serious injury.We currently do not have any information that would warrant further action towards the device manufacturing or devices on the market, however as per our complaint handling processes will continue to monitor the customer experiences with the device for any future information.
|
| |
|
Event or Problem Description
|
|
Manufacturer's reference number: (b)(4).
|
| |
|