MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. SINGLE USE PRELOADED SPHINCTEROTOME V

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. SINGLE USE PRELOADED SPHINCTEROTOME V Back to Search Results
Model Number KD-V611M-07202S
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/21/2023
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported to olympus the single use preloaded sphincterotome v broke while the handle was tightened during an endoscopic retrograde cholangiopancreatography procedure.It was reported that the patient was not under anesthesia at the time of the event.There was no indication of device fragmentation or pieces within patient anatomy.There was no patient injury or medical intervention associated with this event.The event resulted in a ten-minute delay.The procedure was completed using a similar device.
 
Additional Manufacturer Narrative
The device was returned to olympus and is pending device evaluation.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified.A likely mechanism causing the cutting wire breakage might be the following: 1.The cutting wire where the wire coating was torn came into contact with the distal end of the endoscope when the forceps elevator was raised.2.Under circumstances described above 1, an electric conduction was activated.This caused the cutting wire to become hot instantly at the contact point.As a result, the cutting wire broke.A likely factor causing tears of the coated portion of the cutting wire might be the following: 1.It is possible that the slider was slightly pushed causing the cutting wire to deflect.The coated portion of the cutting wire has possibly been rubbed due to the effect of contacting a metal area of the endoscope.However, the exact cause could not be determined.This instruction manual contains the following information: "since the cutting wire is very thin, it may break off in the following cases: the distance between the papilla of vater and the cutting wire is very short, the output is too high or activated while the cutting wire touches metal parts of the endoscope, or the cutting wire is tightened too strong.When the cutting wire breaks off, its proximal end will be retracted toward the endoscope if the slider is pulled.If the slider is pushed, the cutting wire will be pushed out toward the papilla or move sideways.If the cutting wire breaks off, stop the output immediately and pull the slider completely to retract the broken cutting wire into the tube.Then withdraw the sphincterotome from the papilla.Otherwise, patient injury, such as perforations, bleeding, or lacerations within the biliary duct and/or damage of the endoscope could result.¿be sure that the rear end of the cutting wire is extended from the distal end of the endoscope.Insufficient output or unintended tissue injury may occur in case the cutting wire contacts the forceps elevator.¿when inserting the instrument into the endoscope, be sure that the cutting wire is parallel to the tube.Otherwise, the metal part of the forceps elevator may contact the cutting wire and peel off the coating material.¿do not activate output while tissue is in contact with the torn or damaged coated portion of the distal end.If output is activated while tissue is contacting the torn or damaged coated portion due to insertion into or withdrawal from an endoscope, leakage current, decreased output, and/or thermal injury could result.¿if you feel the cutting is blunt, withdraw the sphincterotome from the scope to examine if there is any peel off and tear at the coating portion." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE PRELOADED SPHINCTEROTOME V
Common Device Name
SINGLE USE PRELOADED SPHINCTEROTOME
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18612289
Report Number9614641-2024-00210
Device Sequence Number7668324
Product Code KNS
UDI-Device Identifier04953170403064
UDI-Public04953170403064
Combination Product (Y/N)N
Initial Reporter CountryDA
PMA/510(K) Number
K122505
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 03/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberKD-V611M-07202S
Device Lot Number31V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/30/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/02/2024
Supplement Date Received by Manufacturer03/04/2024
Initial Report FDA Received Date01/31/2024
Supplement Report FDA Received Date03/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/17/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-