Continuation of d10: product id 37642 lot# serial# (b)(6) implanted: explanted: product type programmer, patient section d information references the main component of the system.Other relevant device(s) are: product id: 37642, serial/lot #: (b)(6), ubd: , udi#: (b)(4)h3: analysis of the 37642 patient programmer (ptm) (serial number (b)(6)) revealed a corroded digital board, and corrupt software.This regulatory report is being submitted as part of a retrospective review as part of remediation plan 411.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
It was reported that patient had not used their patient programmer in several months, and when they went to use it, the battery was dead.The caller changed out the batteries and the programmer would not turn on.The patient stated that out of frustration, they had pressed several different buttons.The caller then reported seeing a screen on the programmer saying "system set up" with options of select or exit.Patient services (pss) reviewed meaning of this screen and explained that the programmer would have to be replaced.During the call, the caller noticed acid build up in the battery compartment.No symptoms were reported.
|