| Model Number |
37602 |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3: event date is not known.Please see b5 for approximate date range, if applicable.Continuation of d10: product id: b35300 (b)(6); implant date n/a; explant date: n/a, section d references the main component of the system.Other medical products in use during the event include: brand name ; product id 7482a40 (serial: (b)(6)); product type: 0191-extension; implant date (b)(6) 2008; brand name activa; product id 3387s-40 (lot: v815155); product type: 0200-lead; implant date: (b)(6) 2012.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that patient (pt) in for implantable neurostimulator (ins) change due to routine normal eri.Imped ram in preop showed imped issue on rt ins at 265 ohms for only bipolar pair 1-3x. after replacement to percept new ins impedance issue ran several but saw no change.The impedance issue at 1-3 stayed same.Or current prog does not use those contact pairs.The issue was resolved.No symptoms reported.
|
| |
|
Event or Problem Description
|
|
Further review indicated the event was reported in error.Impedance was 265 ohms which was in the threshold for the device.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: product id 7482a40 lot# serial# (b)(6) implanted: (b)(6) 2008 explanted: product type extension product id 3387s-40 lot# v815155 serial# implanted: (b)(6) 2012 explanted: product type lead product id b35300 lot# serial# (b)(6) implanted: (b)(6) 2024 explanted: product type implantable neurostimulator review of this mdr and/or additional information received shows that there is no information to reasonably suggest that the device in this report may have caused or contributed to a death or serious injury or that the device in this report has malfunctioned.Therefore, this event does not meet the reporting requirements stipulated in 21 cfr 803.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|