U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS Back to Search Results
Model Number 37602
Medical Device Problem Codes Display or Visual Feedback Problem (1184); Failure to Power Up (1476)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Continuation of d10: product id 37642 serial# (b)(6) product type programmer, patient section d information references the main component of the system.Other relevant device(s) are: product id: 37642, serial/lot #: (b)(6), udi#: (b)(4) h3: analysis of the 37642 programmer serial number (b)(6) revealed corrosion.This regulatory report is being submitted as part of a retrospective review as part of remediation plan 411.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the programmer has not worked for a couple of years, the screen is dark and blank despite changing the batteries.Caller checked the battery compartment and reported it looked just fine.Caller said when the batteries are changed, replaced with brand new ones and placed in the battery compartment the correct way it does not power up.They've been using the other one for a couple of years instead and just found out they could get this one replaced.The issue was not resolved.A replacement device will be mailed to the patient.No symptoms were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Common Device Name
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key18614102
Report Number3004209178-2024-02887
Device Sequence Number7565666
Product Code MHY
UDI-Device Identifier00763000100346
UDI-Public00763000100346
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P960009
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/14/2022
Device Model Number37602
Device Catalogue Number37602
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/26/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/20/2022
Initial Report FDA Received Date01/31/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/23/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"SEE H10...."
Patient SexMale
-
-