• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON & CO. (SPARKS) BD PHOENIX¿ AUTOMATED MICROBIOLOGY SYSTEM; SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON & CO. (SPARKS) BD PHOENIX¿ AUTOMATED MICROBIOLOGY SYSTEM; SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION Back to Search Results
Catalog Number 448100
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/05/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
E1: initial reporter phone # (b)(6).The information for the additional 510k is as follows: g5.Pma / 510(k)#: k020321, k020322, k020323, k040099, k131331.H3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported during use of the bd phoenix¿ automated microbiology system there was false resistance in patient results.The results were verified by manual methodology.No health impact or consequence reported.
 
Event or Problem Description
It was reported during use of the bd phoenix¿ automated microbiology system there was false resistance in patient results.The results were verified by manual methodology.No health impact or consequence reported.
 
Additional Manufacturer Narrative
H.6 investigation summary : a results failure was reported on a phoenix 100 instrument.The customer reported a false resistance issue with the instrument.The field service engineer (fse) was dispatched for investigation.The fse verified that the tips used are not elongated, thus potentially causing contamination through the pipette that touches the tube at the time of transfer.The fse carried out the panel preparation process very carefully to avoid contamination and perform sanitization on the pipette.The fse has verified that the false resistance issue has not occurred.This is not true failure of bd instrument.The complaint is unconfirmed.No samples were expected to be received as part of this complaint, and therefore no samples were returned, and no returned material investigation could occur.Dhr review is not required for this complaint as this complaint does not allege an early life failure or failure at installation and the configuration has changed since release from manufacturing due to service repairs/pms.No new risks or hazards, or changes to existing risks/hazards, were identified because of this complaint.Bd quality will continue to monitor for trends associated with the failure mode described in this complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PHOENIX¿ AUTOMATED MICROBIOLOGY SYSTEM
Common Device Name
SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION
Manufacturer (Section D)
BECTON DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
jo doyka
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key18614551
Report Number1119779-2024-00056
Device Sequence Number7668406
Product Code LON
UDI-Device Identifier00382904481001
UDI-Public(01)00382904481001
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
SEE H.10
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number448100
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/05/2024
Supplement Date Received by Manufacturer05/20/2024
Initial Report FDA Received Date01/31/2024
Supplement Report FDA Received Date05/30/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/05/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-