| Model Number |
8864 |
| Medical Device Problem Codes |
Poor Quality Image (1408); Application Program Problem (2880)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/16/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was cancelled.A capital equipment ilab ultrasound imaging system was selected for use.During procedure, a first run was successfully performed.When a second was attempted, no image was displayed.The mdu5+ component seemed to work but the image was black.The procedure was cancelled due to this event.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacture: the device was not returned for analysis; however, a capital equipment technical fse (field service engineer) visited the site to investigate the reported issue with ilab system.Fse reinstalled polaris 2.12 using dvd and performed several runs one after the other successfully.Console worked properly according to boston scientific specifications.
|
| |
|