• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ILAB ULTRASOUND IMAGING SYSTEM; CATHETER, ULTRASOUND, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ILAB ULTRASOUND IMAGING SYSTEM; CATHETER, ULTRASOUND, INTRAVASCULAR Back to Search Results
Model Number 8864
Medical Device Problem Codes Poor Quality Image (1408); Application Program Problem (2880)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/16/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the procedure was cancelled.A capital equipment ilab ultrasound imaging system was selected for use.During procedure, a first run was successfully performed.When a second was attempted, no image was displayed.The mdu5+ component seemed to work but the image was black.The procedure was cancelled due to this event.There were no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by manufacture: the device was not returned for analysis; however, a capital equipment technical fse (field service engineer) visited the site to investigate the reported issue with ilab system.Fse reinstalled polaris 2.12 using dvd and performed several runs one after the other successfully.Console worked properly according to boston scientific specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ILAB ULTRASOUND IMAGING SYSTEM
Common Device Name
CATHETER, ULTRASOUND, INTRAVASCULAR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
150 baytech dr.
san jose CA 95134
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18614851
Report Number2124215-2024-04979
Device Sequence Number7565205
Product Code DQK
UDI-Device Identifier08714729777694
UDI-Public08714729777694
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K151613
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8864
Device Catalogue Number8864
Device Lot Number0000007920
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/16/2023
Initial Report FDA Received Date01/31/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/14/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-