| Model Number |
2102-08-091 |
| Medical Device Problem Code |
Separation Problem (4043)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); Insufficient Information (4580)
|
| Date of Event |
10/30/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The complaint received identified that a fragment of the electrode was left behind in the patient after the electrode was removed.After visual inspection of the returned product, it is clear that the electrode tubing sheared between the third and fourth contacts.This likely occurred due to the force applied when pulling the electrode out.The information provided about the electrode not going in smoothly indicates that the electrode may have been off angle slightly, and because of that, when it was pulled out it may have caught on the anchor bolt which could have caused the electrode to break.
|
| |
|
Event or Problem Description
|
|
Email received from doctor on november 10, 2023.I want to open a "ticket" for the company to review the case we had last week where the electrode got torn.There is a tiny fragment (the tip) that is stuck in the motor cortex so i cannot pull it out.We are still trying to figure out if the patient can have an mri (i believe he can but we are trying to understand) and also what can be the reason.I talked to several friends that are quite experienced with seeg working with pmt electrodes and none had this issue.I myself did at least 50 cases with pmt and more than 500 electrodes and never had this issue.
|
| |
|