| Catalog Number |
14200100 |
| Medical Device Problem Code |
Failure to Align (2522)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As reported: "g4 intermediate nail miss drill was occurred on lateral side during drilling of distal dynamic hole.Pre-calibration showed correct targeting.The cortical bone and nail were in contact at the lateral end, and the physician determined that the miss drill was due to loading on the nail.".
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.Should additional information become available, it will be provided in a supplemental report.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could not be confirmed since the device was not returned for evaluation and no other evidence was provided.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on the available information did the affected devices pass the intra-operative function test before insertion and also the post-operative function test as the instruments are continued to be used.Therefore, a product related issue can be excluded and the event was most likely use related, which is also indicated by statement in the event description that there was possibly to much load on the nail.If more information is provided, the case will be reassessed.
|
| |
|
Event or Problem Description
|
|
As reported: "g4 intermediate nail miss drill was occurred on lateral side during drilling of distal dynamic hole.Pre-calibration showed correct targeting.The cortical bone and nail were in contact at the lateral end, and the physician determined that the miss drill was due to loading on the nail.".
|
| |
|
Search Alerts/Recalls
|