| Model Number |
3300 |
| Medical Device Problem Codes |
Defective Device (2588); Key or Button Unresponsive/not Working (4063)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/19/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for product analysis.Although the device was not returned for analysis, boston scientific performed an investigation with all available information.An unresponsive touchscreen is a known and anticipated potential hazard that can occur during use of the device which is accounted for in device risk documentation and labeling.Boston scientific has received similar reported events where the latitude programmer screen became unresponsive to touch and a complaint review confirmed that the occurrence of harm associated with these events is low.Investigation of this behavior has not found any commonalities (i.E., device serial numbers, date of the event, date of manufacture, etc.) in the reports received.The investigation determined that the cause of touchscreens becoming unresponsive during use was due to the water detection feature within the programmer's liquid crystal display (lcd) touchscreen assembly.Various inputs can induce false positive detections of this water detection feature, which then results in an unresponsive touchscreen.
|
| |
|
Event or Problem Description
|
|
It was reported that while using the pacing system analyzer (psa) programmer function while connected to an implantable device that was in storage mode, the programmer became unresponsive and the programmer screen froze when attempting to toggle out of psa back to the device.The programmer was manually shutdown and restarted, however the screen was still unresponsive.No adverse patient effects were reported, and the programmer was returned for repair.
|
| |
|
Additional Manufacturer Narrative
|
|
This programmer was thoroughly inspected and analyzed upon receipt at our quality assurance laboratory.Visual inspection of the device revealed external damage consistent with the programmer being dropped.The programmer was powered on and it was confirmed that the touchscreen was intermittently unresponsive due to not detecting touch inputs.No error or fault codes were recorded in the programmer's logfile.An unresponsive touchscreen is a known and anticipated potential hazard that can occur during use of the device which is accounted for in device risk documentation and labeling.Boston scientific has received similar reported events where the latitude programmer screen became unresponsive to touch and a complaint review confirmed that the occurrence of harm associated with these events is low.Investigation of this behavior has not found any commonalities (i.E., device serial numbers, date of the event, date of manufacture, etc.) in the reports received.
|
| |
|
Event or Problem Description
|
|
It was reported that while using the pacing system analyzer (psa) programmer function while connected to an implantable device that was in storage mode, the programmer became unresponsive and the programmer screen froze when attempting to toggle out of psa back to the device.The programmer was manually shutdown and restarted, however the screen was still unresponsive.No adverse patient effects were reported, and the programmer was returned for repair.
|
| |
|