• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. BIPAP A40; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. BIPAP A40; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number ESX3100S19
Medical Device Problem Codes Partial Blockage (1065); Contamination (1120); Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/12/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3 other text : device not returned to manufacturer.
 
Event or Problem Description
The manufacturer received information alleging that an inoperative alarm occurred on a ventilator.There was no harm or injury reported.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Additional Manufacturer Narrative
The manufacturer previously reported a ventilator inoperative alarm occurred.There was no harm or injury reported.The device was returned to the manufacturer's product investigation laboratory for investigation.The device's event log was downloaded and reviewed.Although the alarm was not duplicated in the lab, ventilator inoperative alarms were observed in the downloaded event log.Pressure regulation alarms preceded the ventilator inoperative alarms.The device was disassembled, and the etched disk was found to be partially clogged with debris originating external to the device.The device was reassembled but no components were replaced to correct the issue.The device is to be scrapped per the manufacturer's protocol.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP A40
Common Device Name
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key18621089
Report Number2518422-2024-05416
Device Sequence Number12656488
Product Code MNT
UDI-Device Identifier00606959059368
UDI-Public00606959059368
Combination Product (Y/N)N
Initial Reporter CountryPO
PMA/510(K) Number
K121623
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberESX3100S19
Device Catalogue NumberESX3100S19
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/12/2023
Supplement Date Received by Manufacturer08/22/2024
Initial Report FDA Received Date02/01/2024
Supplement Report FDA Received Date08/27/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/24/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-