It was reported that during a shoulder surgery the device did not work optimally.The anchor broke.All broken parts were retrieved from the patient.According to the surgeon, no harm for patient, operator or third party occurred.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause of the reported failure is user error, including improper bone preparation, misaligned insertion, and/or prying or levering the device during insertion.The complaint allegation was not confirmed.The device was not received for evaluation, and no evidence of the failure was provided.
|