| Model Number |
3100 B |
| Medical Device Problem Code |
Increase in Pressure (1491)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Vyaire medical file identification: (b)(4).H10: vyaire field service went onsite replace3 ohms driver.Cannot zero mean pressure order alarm pcb (printed circuit board).
|
| |
|
Event or Problem Description
|
|
It was reported to vyaire medical that the 3100b when device is powered on, the map incrementally increased to max (~60cmh2o) without any intervention.The customer confirmed that there is no patient involvement associated with this reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated information:g3, g6, h2 and h10 additional information:as a resolution, replace pressure transducer pwb assembly.All work accomplished per 3100b service manual.Performed patient circuit calibration and performance check, unit passes.Vent is operational and meets oem (original equipment manufacture) specifications.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: g3, g6, h2, h6 and h11 result of investigation: the vyaire failure analysis laboratory received the suspect component and performed a failure investigation.The reported ¿map incrementally increased to max¿ was not duplicated, however a voltage fluctuation at ambient pressure was found during testing which caused zero calibration reading on test vent to drift outside of the ±0.3cmh2o spec when compared to actual reading from external manometer.
|
| |
|