| Model Number |
N/A |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).H6 - component code - proposed code is mechanical (g04) - provisional bottom.The returned provisional exhibited signs of repeated use (nicked/gouged) and was fractured.Fractured tibial articular surface provisionals (tasps) analyzed by optical microscopy revealed hackle marks, river lines and striations in the case of bending overload and low cycle fatigue culminating in bending overload.The device history records were reviewed and no discrepancies were identified.A definitive root cause could not be determined with the information provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that the tibial articular surface provisional was returned to zimmer biomet in fractured condition.No adverse events were reported as a result of this malfunction.
|
| |
|