| Catalog Number |
CAT01857 |
| Medical Device Problem Code |
Device Contaminated During Manufacture or Shipping (2969)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/12/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was foreign material inside the sterile packaging.
|
| |
|
Event or Problem Description
|
|
It was reported that there was foreign material inside the sterile packaging.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: injector ii needle, cat01857 has a black foreign body inside sterile packaging.I have 2 different lot numbers both with tiny debris inside inner package.Lot# 009723, and 006743.The failure(s) identified in the investigation is consistent with the complaint record.The root cause is a manufacturing issue.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|