| Catalog Number |
13200111 |
| Medical Device Problem Code |
Failure to Align (2522)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/08/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
| |
|
Event or Problem Description
|
|
As reported by the rep: "the surgeon approached me following a short gamma3 nail.He was distally targeting the screws and aimed for the static hole but the screw was placed in the dynamic position and in his words appeared to catch on the edge of the hole.The procedure was completed successfully as the screw went into the dynamic hole position.".
|
| |
|
Event or Problem Description
|
|
As reported by the rep: "the surgeon approached me following a short gamma3 nail.He was distally targeting the screws and aimed for the static hole but the screw was placed in the dynamic position and in his words appeared to catch on the edge of the hole.The procedure was completed successfully as the screw went into the dynamic hole position.".
|
| |
|
Additional Manufacturer Narrative
|
|
Please note correction to section d9/h3 the device was lost by the customer.The reported event could not be confirmed since the device was not returned for evaluation and no other evidence was provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.In case substantive information will become available in future that suggests otherwise we reserve the right to reopen the case.H3 other text : device lost by the customer.
|
| |
|