Catalog Number 13200111
Medical Device Problem Code Failure to Align (2522)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/08/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
As reported by the rep: "the surgeon approached me following a short gamma3 nail.He was distally targeting the screws and aimed for the static hole but the screw was placed in the dynamic position and in his words appeared to catch on the edge of the hole.The procedure was completed successfully as the screw went into the dynamic hole position.".
 
Event or Problem Description
As reported by the rep: "the surgeon approached me following a short gamma3 nail.He was distally targeting the screws and aimed for the static hole but the screw was placed in the dynamic position and in his words appeared to catch on the edge of the hole.The procedure was completed successfully as the screw went into the dynamic hole position.".
 
Additional Manufacturer Narrative
Please note correction to section d9/h3 the device was lost by the customer.The reported event could not be confirmed since the device was not returned for evaluation and no other evidence was provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.In case substantive information will become available in future that suggests otherwise we reserve the right to reopen the case.H3 other text : device lost by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET DEVICE GAMMA3®
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key18623720
Report Number0009610622-2024-00051
Device Sequence Number2406477
Product Code HSB
UDI-Device Identifier04546540716774
UDI-Public04546540716774
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K123401
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number13200111
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/08/2024
Supplement Date Received by Manufacturer03/21/2024
Initial Report FDA Received Date02/01/2024
Supplement Report FDA Received Date04/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-