| Catalog Number |
AB35W06200135 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/18/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that an attempt was made to use a evercross to treat a calcified lesion in the mid superficial femoral artery. degree of tortuosity was little. degree of calcification was moderate.A 6f destination sheath was used. a.035 stiff glide guidewire was used.Embolic protection was not used.An inflation device was used.Device was prepped per ifu.There was a longitudinal balloon burst on the first inflation at 8atms. after loading onto the primary guidewire, the physician noted that contrast was seen exiting the balloon after the tech used an inflation device to take it to nominal.After the burst balloon was removed, the physician asked for a second 6x200 evercross to be opened.After ifu prep and inspection, the physician was able to inflate this pta balloon to nominal pressure and was able to complete procedure with no patient injury reported.
|
| |
|
Search Alerts/Recalls
|
|
|