| Model Number |
26168DM |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the manipulator ring broke off and remained in the vagina.The patient passed the part on the toilet the next day.No negative impact in state of health reported.Nevertheless, the risk of internal injuries and infection must be taken into account as related to this incident.
|
| |
|
Additional Manufacturer Narrative
|
|
The affected device has been requested for investigation by the manufacturer.Device was not yet returned for investigation.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The affected device has been requested for investigation by the manufacturer.Device was returned for investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
During investigation of the returned product (article 26168dm with lot xv01) a break was found at the distal end of the item.The article 26168dm consists of three parts, the anatomical blade of the special screw and the ceramic attachment.The ceramic attachment with the special screw is missing from the examined article.The item was manufactured in october 2014.It is likely that high age and the associated brittle material caused the article to break, possibly there was also incorrect handling which favoured the breakage through excessive force.Internal karl storz reference number: (b)(4).
|
| |
|