BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES BD FACS¿ LYSE WASH ASSISTANT; STATION, PIPETTING DILUTING CLINICAL USE Back to Search Results
Catalog Number 337146
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/30/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D.4.Medical device expiration date: na e.1.Initial reporter address 1: (b)(6).H.3.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported that erroneous results were observed during use with the bd facs¿ lyse wash assistant.The following information was provided by the initial reporter: patient sample checklist: 1.Are there erroneous results on patient samples from diagnostic test? yes 2.Was there any delay of treatment due to the issue? no 3.If patient sample were redrawn, was there any change or delay of treatment? no 4.Was there any physical harm/injury to the patient due to issue? no email: we noticed today that the samples that ran looked a little weird, so we reset them and used the other washer and they then looked normal.Both washers got pm yesterday and i think something happened during that.We did run qc on it this morning and it looked ok but the patient samples looked strange.
 
Additional Manufacturer Narrative
H.6 investigation summary: based on the complaint trend, defect trend, dhr, risk analysis and service activity review, the complaint was confirmed, and the potential cause of the unexpected results was determined to be obstructions in the wash arm.The fse disassembled the arm, cleaned, and lubricated it and subsequent testing showed the instrument was functioning as expected.No one was harmed or injured, and no patients were diagnosed or treated based on any erroneous results.The safety risk of this hazard has been identified to be within the acceptable level.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported that erroneous results were observed during use with the bd facs¿ lyse wash assistant.The following information was provided by the initial reporter: patient sample checklist: 1.Are there erroneous results on patient samples from diagnostic test? yes 2.Was there any delay of treatment due to the issue? no 3.If patient sample were redrawn, was there any change or delay of treatment? no 4.Was there any physical harm/injury to the patient due to issue? no email: we noticed today that the samples that ran looked a little weird, so we reset them and used the other washer and they then looked normal.Both washers got pm yesterday and i think something happened during that.We did run qc on it this morning and it looked ok but the patient samples looked strange.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD FACS¿ LYSE WASH ASSISTANT
Common Device Name
STATION, PIPETTING DILUTING CLINICAL USE
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES
2350 qume drive
san jose CA 95131
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES
2350 qume drive
san jose CA 95131
Manufacturer Contact
fahmy razak - mdr
2350 qume drive
san jose, CA 95131
4089542435
MDR Report Key18624358
Report Number2916837-2024-00012
Device Sequence Number12612008
Product Code JQW
UDI-Device Identifier00382903371464
UDI-Public00382903371464
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number337146
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/30/2024
Supplement Date Received by Manufacturer05/20/2024
Initial Report FDA Received Date02/01/2024
Supplement Report FDA Received Date05/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/30/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-