| Model Number |
DMS22 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
Cerebrospinal Fluid Leakage (1772)
|
| Date of Event |
12/15/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Quality control testing of reserve product met acceptance criteria.
|
| |
|
Event or Problem Description
|
|
Customer reported duramatrix suturable was used on a patient during a chiari procedure on (b)(6) 2023.Customer stated adherus was utilized during the original procedure.A post-op csf leak repair was done on (b)(6) 2023.It was noted that the graft had a hole in the middle of it when revision surgery occurred to repair the leak.Patient status and patient information was requested but not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Quality control testing of reserve product met acceptance criteria for thermal transition and stuture pull out strength.It is expected the product performed as expected.Further trend analysis by clinical affairs and quality assurance concluded there is no identifiable trend and the rate of complaints does not exceed market standards.
|
| |
|