| Medical Device Problem Code |
Computer Software Problem (1112)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/02/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The manufacturer's investigation is on-going and further information will be provided once the investigation has completed.
|
| |
|
Event or Problem Description
|
|
Customer reported that after treatment the recorded beams were not recorded in mosaiq.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d4 corrected, section h4 updated, section h6 updated, section h11 updated.The investigation was completed by conducting a thorough evaluation of the product and the reported information.According to machine and audit logs, a field was entered for treatment verification and the machine transitioned to a ready state.The sdd logs were reviewed and confirmed that the prescribed mu was delivered successfully to the field which the customer already manually recorded.There was no patient mistreatment.Mosaiq did not have any malfunction and was working as designed and intended.For an unknown reason, mosaiq never received data from the linac indicating the beam was being delivered.As a result, mosaiq did not record a treatment for the field.Mosaiq is only able to record according to what it receives from the machine.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: additional information: this is a follow-up to the final report submitted on 07-jun-2024 to notify that due to an administrative error this case was submitted under an incorrect mfr report number (3015232-2024-00007) which captured the wrong manufacturer information.An initial and final (combined) report was submitted under the correct mfr number (see report 2950347-2024-00010) with the correct manufacturer.
|
| |
|
Search Alerts/Recalls
|