• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ELEKTA SOLUTIONS AB MOSAIQ; MOSAIQ ONCOLOGY INFORMATION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ELEKTA SOLUTIONS AB MOSAIQ; MOSAIQ ONCOLOGY INFORMATION SYSTEM Back to Search Results
Medical Device Problem Code Computer Software Problem (1112)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/02/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturer's investigation is on-going and further information will be provided once the investigation has completed.
 
Event or Problem Description
Customer reported that after treatment the recorded beams were not recorded in mosaiq.
 
Additional Manufacturer Narrative
Section d4 corrected, section h4 updated, section h6 updated, section h11 updated.The investigation was completed by conducting a thorough evaluation of the product and the reported information.According to machine and audit logs, a field was entered for treatment verification and the machine transitioned to a ready state.The sdd logs were reviewed and confirmed that the prescribed mu was delivered successfully to the field which the customer already manually recorded.There was no patient mistreatment.Mosaiq did not have any malfunction and was working as designed and intended.For an unknown reason, mosaiq never received data from the linac indicating the beam was being delivered.As a result, mosaiq did not record a treatment for the field.Mosaiq is only able to record according to what it receives from the machine.
 
Additional Manufacturer Narrative
H11: additional information: this is a follow-up to the final report submitted on 07-jun-2024 to notify that due to an administrative error this case was submitted under an incorrect mfr report number (3015232-2024-00007) which captured the wrong manufacturer information.An initial and final (combined) report was submitted under the correct mfr number (see report 2950347-2024-00010) with the correct manufacturer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOSAIQ
Common Device Name
MOSAIQ ONCOLOGY INFORMATION SYSTEM
Manufacturer (Section D)
ELEKTA SOLUTIONS AB
kungstensgatan 18
box 7593
stockholm, SE-10 3 93
SW  SE-103 93
Manufacturer (Section G)
ELEKTA SOLUTIONS AB
400 perimeter center terrace
suite 50
atlanta GA 30346
Manufacturer Contact
cornerstone
london road
crawley, west sussex RH10 -9BL,
MDR Report Key18629738
Report Number3015232-2024-00007
Device Sequence Number18125354
Product Code IYE
UDI-Device Identifier00858164002091
UDI-Public(01)00858164002091(10)2.64.294
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K183034
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/03/2024
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date02/02/2024
Supplement Report FDA Received Date06/07/2024
11/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/24/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-