• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; IMAGINA GASTROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; IMAGINA GASTROSCOPE Back to Search Results
Model Number EG29-I10C
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/29/2024
Event Type  malfunction  
Event Description
The time of event is not during procedure.There was no report of patient harm.Video image failure(objective lens broken).
 
Manufacturer Narrative
This device is classified as import for export, therefore 510k is not applicable.Model eg29-i10c-us is available in the usa with a 510k number k190805.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the objective lens broken.Based on the result, we concluded that it was caused due to the excessive force applied on the objective lens.In addition, our technician confirmed that the bending rubber cut, the segment hard to move, and the cmos module with drive pcb objective lens cracked; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
IMAGINA GASTROSCOPE
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
gurvinder nanda
3 paragon drive
montvale, NJ 07645
2015712318
MDR Report Key18630080
MDR Text Key334743599
Report Number9610877-2024-50974
Device Sequence Number1
Product Code FDS
UDI-Device Identifier04961333237586
UDI-Public04961333237586
Combination Product (y/n)N
Reporter Country CodeKS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEG29-I10C
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/29/2024
Initial Date Manufacturer Received 01/29/2024
Initial Date FDA Received02/02/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-