• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH T-HANDLE WITH QUICK COUPLING; TAP, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH T-HANDLE WITH QUICK COUPLING; TAP, BONE Back to Search Results
Catalog Number 311.44
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/19/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on 19-january-2024 during sterile processing, the instruments in question were found to be broken.Both products have broken locking mechanisms in the handle.There was no reported patient or procedure involvement.No further information is available.This report is for a t-handle with quick coupling.This is report 2 of 2 for (b)(6).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3,h4,h6 the product was not returned to depuy synthes, however photos were provided for review.The photo investigation revealed that 311.44, t-handle with quick coupling the provided evidence was not sufficient to confirm the reported event.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was not confirmed as the observed condition of the 311.44, t-handle with quick coupling would not contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.H4, h6 part:311.44, synthes lot: 4553810, supplier lot: n/a, release to warehouse date: feb 27,2003, manufactured by: synthes brandywine, no non conformance reports were generated during production.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that had the tap holder assembly (locking mechanism) not working as intended, looks like the inner mechanism is broken.The allegation can be confirmed.A dimensional inspection was unable to be performed due to post manufacturing damage.A functional test was performed and the inner mechanism its damage.The device fail the test.The overall complaint was confirmed as the observed condition of the t-handle with quick coupling would contribute to the complained device issue.Based on the investigation findings, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.H3, h6: device history record (dhr) review conducted: part: 311.44, synthes lot: 4553810, supplier lot: n/a, release to warehouse date: feb 27,2003, manufactured by: synthes brandywine, no ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T-HANDLE WITH QUICK COUPLING
Common Device Name
TAP, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES BRANDYWINE
1303 goshen parkway
west chester CO 19380
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key18630275
Report Number8030965-2024-01858
Device Sequence Number18101616
Product Code HWX
UDI-Device Identifier10886982187413
UDI-Public(01)10886982187413
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/27/2023
Device Catalogue Number311.44
Device Lot Number4553810
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/19/2024
Supplement Date Received by Manufacturer02/09/2024
02/21/2024
03/12/2024
04/09/2024
Initial Report FDA Received Date02/02/2024
Supplement Report FDA Received Date02/14/2024
03/05/2024
03/26/2024
04/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/27/2003
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNIVERSAL CHUCK WITH T-HANDLE
Patient SexUnknown
-
-