WILLIAM COOK EUROPE ZENITH ALPHA¿ THORACIC ENDOVASCULAR GRAFT PROXIMAL TAPERED COMPONENTS; MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Catalog Number ZTA-PT-38-34-167-W
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/12/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Description of event according to initial reporter.A patient of undisclosed gender and age underwent a tevar (thoracic endovascular aortic repair) procedure in which the zenith alpha thoracic endovascular graft proximal tapered components, was used.When the graft was unsheathed in the patient the third stent from the top would not open fully.It was constrained about fifty percent.The surgeon used a coda balloon to open the stent successfully.It was a dissection, patient had some tears in their aorta patient outcome: district manager contacted the surgeon one day post-op and surgeon confirmed the patient was doing well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZENITH ALPHA¿ THORACIC ENDOVASCULAR GRAFT PROXIMAL TAPERED COMPONENTS
Common Device Name
MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
MDR Report Key18630473
Report Number3005580113-2024-00011
Device Sequence Number15386365
Product Code MIH
UDI-Device Identifier10827002381800
UDI-Public(01)10827002381800(17)261012(10)E4460006
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/02/2024
Report Date (Section F) 02/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberZTA-PT-38-34-167-W
Device Lot NumberE4460006
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date01/12/2024
Device Age3 MO
Event Location Hospital
Date Report to Manufacturer01/15/2024
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date02/02/2024
Date Device Manufactured10/12/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
Date Report Sent to FDA02/02/2024
-
-