• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. PWRD ECH FLEX 45MM 44CM SHAFT; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. PWRD ECH FLEX 45MM 44CM SHAFT; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number PLE45A
Medical Device Problem Codes Failure to Fire (2610); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/08/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Date sent: 2/2/2024.D4: batch #: unk.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Is the current patient status known? was there any patient consequence or change in the post-operative care of the patient as a result of the event? (extended hospital stay, readmission, re-operation, etc.) a manufacturing record evaluation was performed for the finished ple45a, with lot/batch number a9c67p, and no non-conformances were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during the first stapling, in the making of the y, there was a failure in the stapler, without triggering it.Evidenced a yellow loose part near the safety lock.No patient consequences reported.No further information is available.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 6/5/2024.D4: batch # 272c77.Investigation summary: the product was returned to ethicon endo-surgery for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that one ple45a device was returned with no apparent damage and with a gst45w reload loaded on the device.The reload was received unfired.In addition, the firing trigger switch actuator was received inside of a plastic bag.The device was returned with a non-working firing mechanism.No further functional test could be performed due to the condition of the device.In order to verify the condition of the internal components, the device was disassembled.Upon disassembling, the switch actuator post was found to be broken with the switch actuator loose; which lead to the device not been able to fire.Based on the information currently available, a product issue was identified during the investigation of the sample received.The damage to the actuator post has been correlated to the design.This product issue will be addressed through ethicon endo surgery¿s quality system.Device history review: a manufacturing record evaluation was performed for the finished device batch number 272c77, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PWRD ECH FLEX 45MM 44CM SHAFT
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key18631149
Report Number3005075853-2024-00974
Device Sequence Number15476934
Product Code GDW
UDI-Device Identifier10705036014379
UDI-Public10705036014379
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K110385
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2025
Device Catalogue NumberPLE45A
Device Lot NumberA9C67P
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/08/2024
Supplement Date Received by Manufacturer05/13/2024
Initial Report FDA Received Date02/02/2024
Supplement Report FDA Received Date06/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/28/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-