| Model Number |
6373 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Codes |
Inadequate Pain Relief (2388); Movement Disorder (4412)
|
| Date of Event |
01/09/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Related manufacturer reference number 1627487-2024-00326.It was reported the extensions were replaced due to high impedance and fracture.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event is estimated.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported high impedance was not confirmed.As received, the continuity testing showed the returned both end segments of lead (b) passed isolation resistant testing.The cause of the reported event is due to damage during use.
|
| |
|
Additional Manufacturer Narrative
|
|
H6 - adverse event problem - the clinical codes were corrected to reflect dbs patient symptoms.
|
| |
|