| Model Number |
X SERIES |
| Medical Device Problem Codes |
Failure to Deliver Shock/Stimulation (1133); Delayed Charge Time (2586)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/06/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to defibrillate a 55-year-old male patient, the device took an extended time to charge energy and failed to discharge.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned to zoll medical corporation for evaluation.Instead, the superpower ii battery (sn (b)(6)) was returned for evaluation.The customer's report was verified, however, it was attributed to a depleted battery due to use.A review of the battery log did not observe any discrepancies.The customer was sent a replacement battery.Analysis of reports of this type has not identified an increase in trend.The x series operator's guide states: "always carry at least one fully charged spare battery.If no other source of back-up power is available, two spare batteries are advisable".
|
| |
|