• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT GMBH CALCIUM; AZO DYE, CALCIUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT GMBH CALCIUM; AZO DYE, CALCIUM Back to Search Results
Catalog Number 03L79-22
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/17/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
All available patient information was included.Additional patient details are not available.An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Event or Problem Description
The customer reported a falsely depressed calcium (ca) result for a 41-year-old male patient when running on the architect c16000 processing module.The following data was provided: initial results (mmol/l): ca = 1.3.Repeat results (mmol/l) ca = 2.45.The customer¿s reference ranges were: ca: 2.11-2.52 mmol/l.There was no impact to patient management reported.
 
Additional Manufacturer Narrative
The complaint investigation included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, and device history record review.Return testing was not completed as returns were not available.Data and information provided by the customer were reviewed and support the complaint issue without indication for any additional issue.As part of the troubleshooting the same sample was rerun on the same instrument and same reagent lots and generated higher results.The quality controls were performing within the expected ranges, which shows that the assays were performing as expected.A ticket search by lot number did not identify any increase in complaint activity that was related to the complaint issue.A review of ticket trending was performed, and no trends were identified that were related to the complaint issue.A review of the device history record did not identify any non-conformances or deviations related to the lot number and complaint issue.Labeling was reviewed and adequately addresses the issue under review.Based on the investigation, no systemic issue or product deficiency of the clinical chemistry calcium reagent lot 46188un23 was identified.
 
Event or Problem Description
The customer reported a falsely depressed calcium (ca) result for a 41-year-old male patient when running on the architect c16000 processing module.The following data was provided: initial results (mmol/l): ca = 1.3.Repeat results (mmol/l).Ca = 2.45.The customer¿s reference ranges were: ca: 2.11-2.52 mmol/l.There was no impact to patient management reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CALCIUM
Common Device Name
AZO DYE, CALCIUM
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
Manufacturer (Section G)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM   65205
Manufacturer Contact
nicole jenne
max-planck-ring 2
post market surveillance
wiesbaden 65205
GM   65205
6122582960
MDR Report Key18632135
Report Number3002809144-2024-00027
Device Sequence Number8092168
Product Code CJY
UDI-Device Identifier00380740161491
UDI-Public00380740161491
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K062855
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/01/2024
Device Catalogue Number03L79-22
Device Lot Number46188UN23
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/17/2024
Supplement Date Received by Manufacturer03/05/2024
Initial Report FDA Received Date02/02/2024
Supplement Report FDA Received Date03/07/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/10/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARC C16K PRC MOD, 03L77-01, (B)(6); ARC C16K PRC MOD, 03L77-01, (B)(6)
Patient Age41 YR
Patient SexMale
-
-