|
It was reported that a patient underwent an unknown surgery on (b)(6) 2023 and barbed suture was used.Before use on the patient, when opening the package, it was found that the suture was already detached from the needle.It was a control release needle.Visual inspection was conducted on the returned device.Visual analysis of the returned sample revealed that it was received one needle and suture that pertained to product code vcpb839d.After investigation of the sample, short insertion was observed in the strand.The visual inspection concluded that the combination of the needle/suture was detached from the needle since the insertion end of the suture did not meet the requirement.No adverse patient consequences were reported.Additional information was requested.
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H3 investigation summary: the product was returned for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample revealed that it was received one needle and suture that pertained to product code vcpb839d.After investigation of the sample, short insertion was observed in the strand.The visual inspection concluded that the combination of the needle/suture was detached from the needle since the insertion end of the suture did not meet the requirement.Based on the information currently available, short insertion issue was identified as pull out during the investigation of the sample received.This product issue will be addressed through the ethicon quality system.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.The following information was requested, but unavailable: "what is the lot number? unk.
|