| Catalog Number |
46100-91 |
| Medical Device Problem Codes |
Crack (1135); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/16/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been received for evaluation, however, investigation is not yet complete.
|
| |
|
Event or Problem Description
|
|
The event involved a cath lab w/5 port "off" manifold (600 psi), 72" admin set and 4 ft transpac® iv where the customer reported that a set was leaking and connections were not secure.The customer further stated that the products were cracked, and fluid sprayed out the side of the connection to transducer when connected to pressure and fluid.There was patient involvement but no harm was reported as a consequence of this event.This is report one of two.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported complaint of a leak and connections would not secure was confirmed.During visual inspection, cracks and crazing near the cracks was observed on the female luer on both ends of the manifold.The returned set was primed and pressure leak tested and leaks were observed from the crack of the female luer on both ends of the manifold.The probable cause of the crack and crazing around the crack had occurred due to environmental stress cracking during use.The dhr for lot number 13672865 was reviewed and there were no relevant non-conformances found.
|
| |
|