| Catalog Number |
E2012-10 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/25/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the pruitt f3: carotid shunt balloon is defective on the white end during a left carotid endarterectomy procedure.It was clamped by the surgeon and the shunt was still able to be used.No injury was reported to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.Therefore, we're unable to conclusively determine the root cause of the reported incident.The safety balloon is intended to inflate when excess pressure is applied to the internal carotid balloon to reduce the possibility of injury by over-inflation.It is possible the balloon was over inflated or inflated too rapidly and resulted in the safety balloon inflating instead of the internal carotid balloon.The lot history record for the device was reviewed, no issues were found during manufacturing or packaging that would cause or contribute to the reported event.
|
| |
|