Model Number N/A
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/11/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The manufacturer has received the sample and is pending evaluation.Results are expected soon.
 
Event or Problem Description
It was reported by the customer disconnection of the luer lock from the extension tubing from statlock product.No other information was provided.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Based on a review of this information, the following was concluded: the complaint of a detached luer connector is confirmed; however, the root cause could not be determined.One photograph of an extension set with a y-site and an iv select statlock kit was returned for evaluation.An initial visual observation of the photograph showed an extension set with a y-site.The luer connector was observed to be detached from the distal end of the extension tube.What appears clear fluid in the tube was observed in the returned photograph.Although the detached luer connector was evident in the submitted photograph, inspection of the photograph was insufficient to identify the cause of the damage.Consequently, this complaint is confirmed as ¿cause unknown¿ at this time.This complaint will be recorded for future trending and monitoring purposes.
 
Event or Problem Description
It was reported by the customer disconnection of the luer lock from the extension tubing from statlock product.No other information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STATLOCK IV SELECT MACRO 6IN SMARTSITE Y AND VALVE
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
DAVOL SURGICAL INNOVATIONS -9616067
ave. roberto fierro #6408
parque industrial aeropuerto
cd. juarez, chih s.a. de c.v. 32690
MX   32690
Manufacturer Contact
kayla bevins
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key18634694
Report Number3006260740-2024-00322
Device Sequence Number8092241
Product Code FPA
UDI-Device Identifier00801741082047
UDI-Public(01)00801741082047
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K980992
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/28/2026
Device Model NumberN/A
Device Catalogue NumberIV0564
Device Lot NumberJUHX8593
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/11/2024
Supplement Date Received by Manufacturer03/27/2024
Initial Report FDA Received Date02/02/2024
Supplement Report FDA Received Date04/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-