|
|
| Model Number |
KREULOCK COMPRESSION SCREW, SS, 3.5X12MM |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
01/23/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
On 01/23/2024, it was reported by a sales representative via phone that a patient underwent a clavicle fracture procedure on (b)(6) 2023 where an ar-2652cl clavicle fracture plate, an ar-8835-14 low profile screw, (2) ar-8835-16 low profile screws, an ar-8835cl-12 kreulock compression screw, (2) ar-8835cl-14 kreulock compression screws, an ar-8835-10 low profile screw, an ar-7268 fibertape® cerclage with tigerlink, and an ar-7268t tigertape® cerclage with fiberlink were implanted.Five weeks post-op, the patient reported hearing or feeling a pop.An x-ray was taken that confirmed that the ar-2652cl plate had broken.On (b)(6) 2024, a revision surgery was performed where the implanted devices were removed.The revision was completed using a device from another manufacturer.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint allegation is not confirmed.Upon visual inspection, it was noted that the kreulock compression screw, ss, 3.5x12mm showed no signs of wear or any manufacturing issues.Dimensional testing was performed, and the device met the requirements for drawing c16537-02 in terms of overall length and the outside diameter of the thread length and thread head.
|
| |
|
Search Alerts/Recalls
|
|
|