• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC DE COSTA RICA S.R.L. POLARXFIT; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC DE COSTA RICA S.R.L. POLARXFIT; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Nerve Damage (1979)
Event Date 01/11/2024
Event Type  Injury  
Event Description
It was reported that the patient experienced phrenic nerve damage.During a procedure a polarx catheter was selected for use.The cooling was performed using polarx fit in the order of left superior pulmonary vein, left inferior pulmonary vein, and right inferior pulmonary vein (ripv).However, the ripv ran out of potential at 17 seconds and the movement of the diaphragm weakened; a double stop had to be performed.Twitching was performed several times; however, it did not return to its normal movement; therefore, only the right superior pulmonary vein was isolated with radiofrequency.The procedure was competed without any further complications.The device is not expected to be returned for analysis due to disposal.
 
Manufacturer Narrative
It was indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARXFIT
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
302 parkway
global park, la aurora
heredia
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key18635250
MDR Text Key334492157
Report Number2124215-2024-05731
Device Sequence Number1
Product Code OAE
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/11/2024
Initial Date FDA Received02/02/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-