• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT Back to Search Results
Lot Number 0029984291
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/29/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that stent dislodgement occurred requiring additional intervention.The 90% stenosed target lesion was located in the moderately tortuous and heavily calcified right coronary artery.A 3.50 x 24mm synergy xd drug-eluting stent was advanced for treatment.However, during the procedure, the shaft of the delivery system was compromised and bent resulting to partially deployed the stent.During removal, it dislodged from the system losing the stent in the guide and sheath at the access site.The undeployed stent was removed, and the second stent was placed successfully.No patient complications were reported.
 
Additional Manufacturer Narrative
D4: lot number: updated from 0031289131 to 0029984291.Synergy xd mr us 3.50 x 32mm stent delivery system was returned for analysis.The stent was not returned for analysis.A visual and tactile examination of the hypotube found no issues.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found no issues along the shaft polymer extrusion.The balloon cones were reviewed, and the crimp marks were visible.A review of the manufacturing stent profile data was performed and the stent outer diameter at the time of manufacture was within max crimped stent profile measurement as per specification.The bumper tip showed no signs of distal tip damage.The reported kinks were not identified during the analysis.No other device issues were identified during the returned product analysis.
 
Event or Problem Description
It was reported that stent dislodgement occurred requiring additional intervention.The 90% stenosed target lesion was located in the moderately tortuous and heavily calcified right coronary artery.A 3.50 x 24mm synergy xd drug-eluting stent was advanced for treatment.However, during the procedure, the shaft of the delivery system was compromised and bent resulting to partially deployed the stent.During removal, it dislodged from the system losing the stent in the guide and sheath at the access site.The undeployed stent was removed, and the second stent was placed successfully.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18635764
Report Number2124215-2024-04205
Device Sequence Number7954731
Product Code NIQ
UDI-Device Identifier08714729981084
UDI-Public08714729981084
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/03/2024
Device Lot Number0029984291
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer02/08/2024
Initial Report FDA Received Date02/02/2024
Supplement Report FDA Received Date03/01/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/04/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-