MMJ SA DE CV (USD) MON-A-THERM; THERMOMETER, ELECTRONIC, CLINICAL Back to Search Results
Model Number 90050
Medical Device Problem Codes Device Damaged Prior to Use (2284); High Readings (2459); Incomplete or Inadequate Connection (4037)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D10 concomitant product: 90050, 90050 400 general purpose sensor 9 x50 (lot#21b0689jzx).H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the product produced more heat than expected, the components failed to connect correctly and there was damage to the unit.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: do not use if package is opened or damaged.Use only with compatible interconnect cables and monitors.Not doing so may result in an inaccurate measured temperature.Do not modify this device.Modifications could result in inaccurate temperature measurement.Do not attempt to force connection of the sensor.The connectors are keyed by shape.Forcing the connection could result in damage to the sensor or cable connector and a resulting inability to function.Ensure that the cable connector does not lie in an area where solutions may pool during the surgical procedure.This report is being submitted as part of a retrospective review and remediation per crp no.1000199834/nc36.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a procedure, the device had temperature results issue and a damage connector.The thermal probe fi rst recorded a temperature of 34.1 degrees in the patient and then the temperature rose sharply (in less than 2 minutes) to 40.5 degrees suspecting a potential malignant hyperthermia in the patient.After control by tympanic thermometer, the temperature was 35.1 degrees in each ear.The thermal probe was then removed. unfortunately, this is not the first time that this type of thermal probe has found erroneous data.  in addition, when removing the thermal probe, the connector separated from the rest of the thermal probe and got stuck on the cable connecting to the scope (as with each removal) complicating the change of thermal probe.It was then not possible to put back a thermal probe, the thermal monitoring was interrupted for this patient.There was no reported patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MON-A-THERM
Common Device Name
THERMOMETER, ELECTRONIC, CLINICAL
Manufacturer (Section D)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX  32590
Manufacturer (Section G)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX   32590
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18636022
Report Number2936999-2024-00220
Device Sequence Number12785483
Product Code FLL
UDI-Device Identifier20884521031606
UDI-Public20884521031606
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K874514
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/28/2026
Device Model Number90050
Device Catalogue Number90050
Device Lot Number21A0886JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/09/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/05/2022
Initial Report FDA Received Date02/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PLEASE SEE NOTE ON H10
Patient SexUnknown
-
-