• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS ULTRASOUND, INC EPIQ CVX; SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS ULTRASOUND, INC EPIQ CVX; SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC Back to Search Results
Catalog Number 795231
Medical Device Problem Code Unintended System Motion (1430)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/25/2024
Type of Reportable Event Malfunction
Event or Problem Description
A customer reported the swivel mechanism of their epiq elite ultrasound system¿s control panel did not lock properly while transporting the unit within the user facility.The failure occurred outside of clinical use, and no patient or user was harmed as a result of the issue.The philips service engineer repaired the system by reseating and securing the bushing.A thorough investigation was performed to identify the root cause of the reported issue.The engineering team determined the failure was caused by a hardware issue in the articulating arm's latching mechanism, which prevented the locking solenoid from fully engaging.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIQ CVX
Common Device Name
SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC
Manufacturer (Section D)
PHILIPS ULTRASOUND, INC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS ULTRASOUND, INC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
thuy nguyen
22100 bothell everett highway
bothell, WA 98021
MDR Report Key18636230
Report Number3019216-2024-00014
Device Sequence Number20774247
Product Code IYN
UDI-Device Identifier00884838097933
UDI-Public00884838097933
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K132304
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number795231
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/25/2024
Initial Report FDA Received Date02/02/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-