AOMORI OLYMPUS CO., LTD. DISPOSABLE DISTAL ATTACHMENT Back to Search Results
Model Number D-201-11804
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Abscess (1690); Hemorrhage/Blood Loss/Bleeding (1888); Bowel Perforation (2668)
Date of Event 05/24/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The suspect device has not been returned to olympus for evaluation.The investigation is in process.The literature article is attached for additional information.Once the investigation has been completed, a supplemental report will be submitted with device evaluation results.
 
Event or Problem Description
Olympus reviewed the following literature titled "usefulness of the s-o clip for duodenal endoscopic submucosal dissection: a propensity score-matched study." this retrospective study evaluated the efficacy and safety of duodenal esd using an s-o clip for snadets.The retrospective study was patients who underwent esd for snadets between january 2011 and december 2021.Propensity score matching analysis was used to compare patients who underwent duodenal esd with the s-o clip (s-o group) and those who underwent conventional esd (control group).Sixty patients with snadets who were treated with esd were enrolled.After propensity score matching, 16 pairs were created: 43 and 17 in the s-o and control groups, respectively.All esd procedures were performed using a single-channel endoscope (gif-h260j) with a transparent hood (d-201-11804) and hook knife (kd-625lr).S-o group: 43 patients delayed bleeding: 2 patients control group: 17 patients intraoperative perforation: 4 patients delayed perforation: 1 patient in case of delayed complications, computed tomography and blood tests were performed.Delayed bleeding treated with endoscopic hemostasis after esd.Most of the cases of intraoperative perforation were successfully managed with conservative treatment alone.However, some cases required conversion to surgery.Another patient in the study developed a retroperitoneal abscess due to an intraoperative perforation and was hospitalized for 67 days.This literature article requires 3 reports.The related patient identifiers are as follows: (b)(6): gif-q260j (b)(6): kd-625lr (b)(6): d-201-11804 this medwatch report is for patient identifier (b)(6).There is no report of any olympus device malfunction in any procedure described in this study.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record was unable to be reviewed for this device since the serial and/or lot number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, the relationship between the device and the adverse event cannot be confirmed.There was no complaint reported on the subject device.There is no evidence of an olympus device malfunction.Therefore, the root cause cannot be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISPOSABLE DISTAL ATTACHMENT
Common Device Name
DISPOSABLE DISTAL ATTACHMENT
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18636449
Report Number9614641-2024-00243
Device Sequence Number7763426
Product Code FDS
UDI-Device Identifier04953170453854
UDI-Public04953170453854
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K984358
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Study,Literature,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberD-201-11804
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/08/2024
Supplement Date Received by Manufacturer03/15/2024
Initial Report FDA Received Date02/02/2024
Supplement Report FDA Received Date03/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GIF-Q260J; KD-625LR
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-