| Catalog Number |
9-ASD-038 |
| Medical Device Problem Code |
Difficult to Fold, Unfold or Collapse (1254)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/10/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that a 38mm amplatzer septal occluder was selected for implant on 10 (b)(6) 2024.During implant the device took on a cobra shape.The device was removed from the patient and replaced with a new non-abbott device.There were no interactions with cardiac structures.There were no angulations or kinks noted in the delivery system.The patient remained hemodynamically stable throughout the procedure.There were no clinically significant delays in the procedure.There were no adverse effects to the patient.The patient status was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of device deformity could not be confirmed.The device was returned to abbott for investigation and the device met functional specifications when analyzed under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.Information from the field indicated that there was no interaction with cardiac structure and no any angulation kink noticed in the delivery system, and unknown size for the delivery system was used.Based on the available information, the cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
|
| |
|
Search Alerts/Recalls
|