| Model Number |
M0061921330 |
| Medical Device Problem Codes |
Break (1069); Deformation Due to Compressive Stress (2889)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/10/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0401 captures the reportable event of stent shaft broken.Imdrf device code a040601 captures the reportable event of stent buckled material, inside the patent.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a polaris ultra ureteral stent was used to treat kidney stones during a ureteroscopy procedure in the urinary tract, uterus and urethra.The procedure date was nor reported.During the procedure, there was a device malfunction.It was noted that the stent was found broken.The procedure was successfully completed with another polaris ultra ureteral stent.There were no patient complications reported as a result of this event.A photo of the complaint device was provided and showed that the stent shaft broken and buckled.
|
| |
|