| Model Number |
2464MO-20 |
| Medical Device Problem Code |
Disconnection (1171)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/12/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported to vyaire medical that they are currently taping down the 2464aao-20 - leg attachment pad for corometrics fetal scalp electrode cable to the leg plate so that it will stay attached.The customer confirmed that there was no patient harm associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: 81 other - at this time, the suspect device has not been returned for evaluation.No root cause has been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported to vyaire medical that they are currently taping down the 2464aao-20 - leg attachment pad for corometrics fetal scalp electrode cable to the leg plate so that it will stay attached.The customer confirmed that there was no patient harm associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Customer complained about leg attachment pads.It was confirmed, that changing the fse cable fixed the issue.The photo provided confirmed that the cable was of older revision, from previous supplier.Reviewed sha - rmf-hl-0060 - ob monitoring products - fse cables - rev06 - (b)(6) 2024.Closest risk could be sha-hl-0060-c-018 loss of function (e.G.Unexpected loss of mechanical or electrical integrity, fatigue failure), most probable cause failure of reusable product due to repeated and extensive use of product leading to loose connector fit.
|
| |
|