DEPUY ORTHOPAEDICS INC US PINN MAR +4 10D 28IDX46OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS Back to Search Results
Catalog Number 121928146
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Foreign Body Reaction (1868)
Date of Event 01/25/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Patient was revised on (b)(6)2022 for unknown reason, and later re-revised on (b)(6)2023, where the head and insert were replaced due to metallosis.No surgical delay reported.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d9 if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint#: (b)(4).Investigation summary: according to the information received, description: 1.Surgery with novelty since reference cup implant batch is removed, acetabular screws length 20 ref, lot, acetabular screw length 25 ref, lot, reference insert, lot, ceramic head reference, lot, this patient has had several interventions the cup was placed on (date of implantation removed) and was re-operated where the insect and head are placed on the day (date of re-intervention) and today the reintervention is done again because the material generated metallosis and the decision is made to place the cemented cup and the disc insert and head 222,225 +7 johnson , the extracted material is left for study and delivered to the stock manager and the head of centra, photo evidence is sent.The product was not returned to depuy synthes, however photos were provided for review.The photo investigation revealed that several anti rotation devices have been sheared off from the liner.Additionally, the rim of the liner presents deformation.Based on the observed damage on the liner and the condition of the cup and the head.It is reasonable to conclude that the liner disassociated from the cup have causing this defect.There is no indication that a design or manufacturing issue has caused or contributed to the reported event.Although the liner failure cannot be traced to design or manufacturing, a definite root cause cannot be established due to there being multiple factors that may influence eccentric loading.Consideration must be given to all potential influences such as surgical process, patient variables (e.G.Activity level and use), anatomical considerations and patient changes over time.Any conclusions from the investigational input provided have to be placed into context with all other relevant factors.It is highly recommended to the patient to consult with their healthcare professional for further assessment.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was confirmed as the observed condition of the liner would contribute to the complained device issue.Based on the investigation findings, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing record evaluation was performed for the finished device m2176a/121928146 number, and no non-conformance's /manufacturing irregularities were identified.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : description: 1.Surgery with novelty since reference cup implant batch is removed, acetabular screws length 20 ref lot, acetabular screw length 25 ref * lot *, reference insert * lot *, ceramic head reference * lot *, this patient has had several interventions the cup was placed on (date of implantation removed) and was re-operated where the insect and head are placed on the day (date of re-intervention) and today the reintervention is done again because the material generated metallosis and the decision is made to place the cemented cup and the disc insert and head 222,225 +7 johnson , the extracted material is left for study and delivered to the stock manager and the head of centra, photo evidence is sent.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual inspection of the returned sample revealed that several anti rotational devices have been sheared off from the pinn mar +4 10d 28idx46od.Additionally, the rim of the liner present deformation.Based on the observed damage on the liner it is reasonable to conclude that the liner disassociated from the cup.Also the liner exhibits signs of broken but it could be due for the extraction process.There is no indication that a design or manufacturing issue has caused or contributed to the reported event.Although the liner failure cannot be traced to design or manufacturing, a definite root cause cannot be established due to there being multiple factors that may influence eccentric loading.Consideration must be given to all potential influences such as surgical process, patient variables (e.G.Activity level and use), anatomical considerations and patient changes over time.Any conclusions from the investigational input provided have to be placed into context with all other relevant factors.It is highly recommended to the patient to consult with their healthcare professional for further assessment.A functional test was not performed as it is not applicable to the complaint condition.A dimensional inspection was unable to be performed due to post manufacturing damage.The overall complaint was confirmed as the observed condition of the pinn mar +4 10d 28idx46od would contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : a manufacturing record evaluation was performed for the finished device m2176a/121928146 number, and no non-conformances /manufacturing irregularities were identified.Note: dhr information was retrieved from (b)(4) as it is the same lot number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN MAR +4 10D 28IDX46OD
Common Device Name
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY RAYNHAM, A DIV. OF DEPUY ORTHO 1219655
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key18637056
Report Number1818910-2024-02287
Device Sequence Number7930983
Product Code LPH
UDI-Device Identifier10603295013440
UDI-Public10603295013440
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K033273
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 02/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number121928146
Device Lot NumberM2176A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/25/2024
Supplement Date Received by Manufacturer02/01/2024
02/07/2024
02/13/2024
04/04/2024
04/24/2024
04/25/2024
Initial Report FDA Received Date02/03/2024
Supplement Report FDA Received Date02/06/2024
02/13/2024
02/15/2024
04/05/2024
04/25/2024
04/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/17/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DELTA CER HEAD 12/14 28MM +1.5; PINN CAN BONE SCREW 6.5MMX20MM; PINN CAN BONE SCREW 6.5MMX25MM; PINNACLE BANTAM ACET CUP 46MM
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-